Realistic Total-Body J-PET Geometry Optimization--Monte Carlo Study
J. Baran, W. Krzemień, L. Raczyński, M. Bała, A. Coussat, S. Parzych, N. Chug, E. Czerwiński, C. Oana Curceanu, M. Dadgar, K. Dulski, K. Eliyan, J. Gajewski, A. Gajos, B. Hiesmayr, K. Kacprzak, Ł. Kapłon, K. Klimaszewski, G. Korcyl, T. Kozik, D. Kumar, S. Niedźwiecki, D. Panek, E. Perez del Rio, A. Ruciński, S. Sharma, Shivani, R.Y. Shopa, M. Skurzok, E. Stępień, F. Tayefiardebili, K. Tayefiardebili, W. Wiślicki, P. Moskal
abstract
Total-Body PET imaging is one of the most promising newly introduced modalities in the medical diagnostics. State-of-the-art PET scanners use inorganic scintillators such as L(Y)SO or BGO, however, those technologies are very expensive, prohibitng the broad total-body PET applications. We present the comparative studies of performance characteristics of the cost-effective Total-Body PET scanners using Jagiellonian PET (J-PET) technology that is based on plastic scintillators. Here, we investigated in silico five realistic Total-Body scanner geometries, varying the number of rings, scanner radius, and distance between the neighbouring rings. Monte Carlo simulations of two NEMA phantoms (2-meter sensitivity line source and image quality) and the anthropomorphic XCAT phantom, were used to assess the performance of the tested geometries. We compared the sensitivity profiles and we performed the quantitative analysis of the reconstructed images by using the quality metrics such as contrast recovery coefficient, background variability and root mean squared error. The optimal scanner design was selected for the first Total-Body J-PET scanner configuration.
Feasibility of the J-PET to monitor range oftherapeutic proton beams
J. Baran, D. Borys, K. Brzeziński, J. Gajewski, M. Silarski, N. Chug, A. Coussat, E. Czerwiński, M. Dadgar, K. Dulski, K.V. Eliyan, A. Gajos, K. Kacprzak, Ł. Kapłon, K. Klimaszewski, P. Konieczka, R. Kopeć, G. Korcyl, T. Kozik, W. Krzemień, D. Kumar, A.J. Lomax, K. McNamara, S. Niedźwiecki, P. Olko, D. Panek, S. Parzych, E. Perez del Rio, L. Raczyński, M. Simbarashe, S. Sharma, Shivani, R.Y. Shopa, T. Skóra, M. Skurzok, P. Stasica, E.Ł. Stępień, K. Tayefi, F. Tayefi, D.C. Weber, C. Winterhalter, W. Wiślicki, P. Moskal, A. Ruciński
abstract
Objective: The aim of this work is to investigate the feasibility of the JagiellonianPositron Emission Tomography (J-PET) scanner for intra-treatment proton beamrange monitoring. Approach: The Monte Carlo simulation studies with GATE and PET imagereconstruction with CASToR were performed in order to compare six J-PET scannergeometries (three dual-heads and three cylindrical). We simulated proton irradiationof a PMMA phantom with a Single Pencil Beam (SPB) and Spread-Out BraggPeak (SOBP) of various ranges. The sensitivity and precision of each scanner werecalculated, and considering the setup?s cost-effectiveness, we indicated potentiallyoptimal geometries for the J-PET scanner prototype dedicated to the proton beamrange assessment. Main results: The investigations indicate that the double-layer cylindrical andtriple-layer double-head configurations are the most promising for clinical application.We found that the scanner sensitivity is of the order of 10?5coincidences per primaryproton, while the precision of the range assessment for both SPB and SOBP irradiationplans was found below 1 mm. Among the scanners with the same number of detectormodules, the best results are found for the triple-layer dual-head geometry. Significance: We performed simulation studies demonstrating that the feasibilityof the J-PET detector for PET-based proton beam therapy range monitoring ispossible with reasonable sensitivity and precision enabling its pre-clinical tests in theclinical proton therapy environment. Considering the sensitivity, precision and cost-effectiveness, the double-layer cylindrical and triple-layer dual-head J-PET geometryconfigurations seem promising for the future clinical application. Experimental testsare needed to confirm these findings.
J-PET application as a Compton camera for proton beam range verification: A preliminary study
M. K. Kozani, A. Rucinski, P. Moskal
abstract
Hybrid in-beam PET/Compton camera imaging currently shows a promising approach to use of the quasi-real-time range verification technique in proton therapy. This work aims to assess the capability of utilizing a configuration of the Jagiellonian-positron emission tomography (J-PET) scanner made of plastic scintillator strips, so as to serve as a Compton camera for proton beam range verification. This work reports the production yield results obtained from the GATE/Geant4 simulations, focusing on an energy spectrum (4.2?4.6) MeV of prompt gamma (PG) produced from a clinical proton beam impinging on a water phantom. To investigate the feasibility of J-PET as a Compton camera, a geometrical optimisation was performed. This optimisation was conducted by a point spread function (PSF) study of an isotropic 4.44 MeV gamma source. Realistic statistics of 4.44 MeV PGs obtained from the prior step were employed, simulating interactions with the detector. A sufficient number of detected photons was obtained for the source position reconstruction after performing a geometry optimisation for the proposed J-PET detector. Furthermore, it was demonstrated that more precise calculation of the total deposited energy of coincident events plays a key role in improving the image quality of source distribution determination. A reasonable spatial resolution of 6.5 mm FWHM along the actual proton beam direction was achieved for the first imaging tests. This preliminary study has shown notable potential in using the J-PET application for in-beam PET/Compton camera imaging at quasi-real-time proton range monitoring in future clinical use.
Comparison of cell casted and 3D-printed plastic scintillators for dosimetry applications
D. Kulig, Ł. Kapłon, G. Moskal, S. Beddar, T. Fiutowski, W. Górska, J. Hajduga, P. Jurgielewicz, D. Kabat, K. Kalecińska, M. Kopeć, S. Koperny, B. Mindur, J. Moroń, S. Niedźwiecki, M. Silarski, F. Sobczuk, T. Szumlak, A. Ruciński
abstract
Currently, the most used methods of plastic scintillator (PS) manufacturing are cell casting and bulk polymerisation, extrusion, injection molding, whereas digital light processing (DLP) 3D printing technique has been recently introduced. For our research, we measured blue-emitting EJ-200, EJ-208, green-emitting EJ-260, EJ-262 cell cast and two types of blue-emitting DLP-printed PSs. The light output of the samples, with the same dimension of 10 mm × 10 mm × 10 mm, was compared. The light output of the samples, relative to the reference EJ-200 cell-cast scintillator, equals about 40?49 and 70?73% for two types of 3D-printed, and two green-emitting cell-casted PSs, respectively. Performance of the investigated scintillators is sufficient to use them in a plastic scintillation dosemeter operating in high fluence gamma radiation fields.
Detection of range shifts in proton beam therapy using the J-PET scanner: a patient simulation study
K. Brzeziński, J. Baran, D. Borys, J. Gajewski, N. Chug, A. Coussat, E. Czerwiński, M. Dadgar, K. Dulski, K.V. Eliyan, A. Gajos, K. Kacprzak, Ł. Kapłon, K. Klimaszewski, P. Konieczka, R. Kopeć, G. Korcyl, T. Kozik, W. Krzemień, D. Kumar, A.J. Lomax, K. McNamara, S. Niedźwiecki, P. Olko, D. Panek, S. Parzych, E. Perez del Rio, L. Raczyński, S. Sharma, Shivani, R.Y. Shopa, T. Skóra, M. Skurzok, P. Stasica, E.Ł. Stępień, K. Tayefi, F. Tayefi, D.C. Weber, C. Winterhalter, W. Wiślicki, P. Moskal, A. Ruciński
abstract
Objective. The Jagiellonian positron emission tomography (J-PET) technology, based on plastic scintillators, has been proposed as a cost effective tool for detecting range deviations during proton therapy. This study investigates the feasibility of using J-PET for range monitoring by means of a detailed Monte Carlo simulation study of 95 patients who underwent proton therapy at the Cyclotron Centre Bronowice (CCB) in Krakow, Poland. Approach. Discrepancies between prescribed and delivered treatments were artificially introduced in the simulations by means of shifts in patient positioning and in the Hounsfield unit to the relative proton stopping power calibration curve. A dual-layer, cylindrical J-PET geometry was simulated in an in-room monitoring scenario and a triple-layer, dual-head geometry in an in-beam protocol. The distribution of range shifts in reconstructed PET activity was visualized in the beam's eye view. Linear prediction models were constructed from all patients in the cohort, using the mean shift in reconstructed PET activity as a predictor of the mean proton range deviation. Main results. Maps of deviations in the range of reconstructed PET distributions showed agreement with those of deviations in dose range in most patients. The linear prediction model showed a good fit, with coefficient of determination r2 = 0.84 (in-room) and 0.75 (in-beam). Residual standard error was below 1 mm: 0.33 mm (in-room) and 0.23 mm (in-beam). Significance. The precision of the proposed prediction models shows the sensitivity of the proposed J-PET scanners to shifts in proton range for a wide range of clinical treatment plans. Furthermore, it motivates the use of such models as a tool for predicting proton range deviations and opens up new prospects for investigations into the use of intra-treatment PET images for predicting clinical metrics that aid in the assessment of the quality of delivered treatment.
Comparative studies of plastic scintillator strips with high technical attenuation length for the total-body J-PET scanner
Ł. Kapłon, J. Baran, N. Chug, A. Coussat, C. Curceanu, E. Czerwiński, M. Dadgar, K. Dulski, J. Gajewski, A. Gajos, B. Hiesmayr, E. Kavya Valsan, K. Klimaszewski, G. Korcyl, T. Kozik, W. Krzemień, D. Kumar, G. Moskal, S. Niedźwiecki, D. Panek, S. Parzych, E. Pérez del Rio, L. Raczyński, A. Ruciński, S. Sharma, S. Shivani, R. Shopa, M. Silarski, M. Skurzok, E. Stępień, F. Tayefi Ardebili, K. Tayefi Ardebili, W. Wiślicki, P. Moskal
abstract
Plastic scintillator strips are considered as one of the promising solutions for the cost-effective construction of total-body positron emission tomography, (PET) system. The purpose of the performed measurements is to compare the transparency of long plastic scintillators with dimensions 6 mm x 24 mm x 1000 mm and with all surfaces polished. Six different types of commercial, general purpose, blue-emitting plastic scintillators with low attenuation of visible light were tested, namely: polyvinyl toluene-based BC-408, EJ-200, RP-408, and polystyrene-based Epic, SP32 and UPS-923A. For determination of the best type of plastic scintillator for total-body Jagiellonian positron emission tomograph (TB-J-PET) construction, emission and transmission spectra, and technical attenuation length (TAL) of blue light-emitting by the scintillators were measured and compared. The TAL values were determined with the use of UV lamp as excitation source, and photodiode as light detector. Emission spectra of investigated scintillators have maxima in the range from 420 nm to 429 nm. The BC-408 and EJ-200 have the highest transmittance values of about 90% at the maximum emission wavelength measured through a 6 mm thick scintillator strip and the highest technical attenuation length reaching about 2000 mm, allowing assembly of long detection modules for time-of-flight (TOF) J-PET scanners. Influence of the 6 mm × 6 mm, 12 mm × 6 mm, 24 mm × 6 mm cross-sections of the 1000 mm long EJ-200 plastic scintillator on the TAL and signal intensity was measured. The highest TAL value was determined for samples with 24 mm × 6 mm cross-section.
TOF MLEM Adaptation for the Total-Body J-PET with a Realistic Analytical System Response Matrix
R.Y. Shopa, J. Baran, K. Klimaszewski, W. Krzemień, L. Raczyński, W. Wiślicki, K. Brzeziński, N. Chug, A. Coussat, C. Curceanu, E. Czerwiński, M. Dadgar, K. Dulski, J. Gajewski, A. Gajos, B.C. Hiesmayr, E. Kavya Valsan, G. Korcyl, T. Kozik, D. Kumar, Ł. Kapłon, G. Moskal, S. Niedźwiecki, D. Panek, S. Parzych, E. Pérez del Rio, A. Ruciński, S. Sharma, Shivani, M. Silarski, M. Skurzok, E. Stepień, F. Tayefi Ardebili, K. Tayefi Ardebili, P. Moskal
abstract
We report a study of the original image reconstruction algorithm based on the time-of-flight maximum likelihood expectation maximisation (TOF MLEM), developed for the total-body (TB) Jagiellonian PET (J-PET) scanners. The method is applicable to generic cylindrical or modular multi-layer layouts and is extendable to multi-photon imaging. The system response matrix (SRM) is represented as a set of analytical functions, uniquely defined for each pair of plastic scintillator strips used for the detection. A realistic resolution model (RM) in detector space is derived from fitting the Monte Carlo simulated emissions and detections of annihilation photons on oblique transverse planes. Additional kernels embedded in SRM account for TOF, parallax effect and axial smearing. The algorithm was tested on datasets, simulated in GATE for the NEMA IEC and static XCAT phantoms inside a 24-module 2-layer TB J-PET. Compared to the reference TOF MLEM with none or a shift-invariant RM, an improvement was observed, as evaluated by the analysis of image quality, difference images and ground truth metrics. We also reconstructed the data with additive contributions, pre-filtered geometrically and with non-TOF scatter correction applied. Despite some deterioration, the obtained results still capitalise on the realistic RM with better edge preservation and superior ground truth metrics. The envisioned prospects of the TOF MLEM with analytical SRM include its application in multi-photon imaging and further upgrade to account for the non-collinearity, positron range and other factors.
A modular data acquisition system for reconstruction of radiation dose spatial distribution in radiotherapy treatment planning
P. Jurgielewicz, M. Filipek, T. Fiutowski, D. Kabat, K. Kalecińska, Ł. Kapłon, M. Kopeć, S. Koperny, D. Kulig, J. Moroń, G. Moskal, A. Ruciński, P. Wiącek, T. Szumlak, B. Mindur
abstract
In this work we propose the complete Data Acquisition (DAQ) system for measurement of volumetric radiotherapeutic dose deposition in tissue-like phantoms based on 3D printed plastic scintillators. The DAQ is easily extensible thanks to the modular architecture of its hardware and software components. We show results from the full measurement chain indicating proper operation of the system.
A reconfigurable detector for measuring the spatial distribution of radiation dose for applications in the preparation of individual patient treatment plans
M. Kopeć, T. Fiutowski, P. Jurgielewicz, D. Kabat, K. Kalecińska, Ł. Kapłon, S. Koperny, D. Kulig, J. Moroń, G. Moskal, A. Ruciński, P. Wiącek, T. Szumlak, B. Mindur
abstract
In this work, a novel reconfigurable Dose-3D detector intended for a full spatial therapeutic dose measurement to improve radiotherapy treatment planning is presented. The device is composed of a reconfigurable detection phantom allowing patient-centric adjustments to its geometry, a scalable data acquisition system (including hardware, firmware, and low-level software) designed to change with the phantom?s configuration seamlessly, and a high-level software package for tumour geometry extraction based on computer tomography scans. Extracted geometry will be used in the Monte Carlo simulations and the configuration of the phantom. Each of the components to be used in the measurement system has been assessed obtaining the following results. The scintillating voxels? light output is sufficient. The data acquisition system with its hardware and software has been tested using artificial testing signals and laser light proving a reliable and robust means of physics data reconstruction.
A reconfigurable detector for measuring the spatial distribution of radiation dose for applications in the preparation of individual patient treatment plans
M. Kopeć, T. Fiutowski, P. Jurgielewicz, D. Kabat, K. Kalecińska, Ł. Kapłon, S. Koperny, D. Kulig, J. Moroń, G. Moskal, A. Ruciński, P. Wiącek, T. Szumlak, B. Mindur
abstract
In this work, a novel reconfigurable Dose-3D detector intended for a full spatial therapeutic dose measurement to improve radiotherapy treatment planning is presented. The device is composed of a reconfigurable detection phantom allowing patient-centric adjustments to its geometry, a scalable data acquisition system (including hardware, firmware, and low-level software) designed to change with the phantom?s configuration seamlessly, and a high-level software package for tumour geometry extraction based on computer tomography scans. Extracted geometry will be used in the Monte Carlo simulations and the configuration of the phantom. Each of the components to be used in the measurement system has been assessed obtaining the following results. The scintillating voxels? light output is sufficient. The data acquisition system with its hardware and software has been tested using artificial testing signals and laser light proving a reliable and robust means of physics data reconstruction.
Medical imaging data analysis using 3D deep learning models towards improving the individual treatment plans
K. Kalecińska, T. Fiutowski, P. Jurgielewicz, D. Kabat, B. Rachwał, Ł. Kapłon, M. Kopeć, S. Koperny, D. Kulig, J. Moroń, G. Moskal, A. Ruciński, P. Wiącek, B. Mindur, T. Szumlak
abstract
This work is a part of a research project aiming at delivering the next generation active medical phantom, Dose-3D, with high spatial granulation for quasi-real time measurement of the volumetric radiotherapeutic dose deposited during photon therapy. The preliminary results, discussed here, pertain to the intelligent medical data augmentation using Generative Adversarial Networks (GANs) technique implemented inside MONAI framework. However, in the scope of the project, we perform a broad search for the most efficient and advanced Deep Learning (DL) models to create tools for 3D Computed Tomography (CT) images segmentation and cancer diagnosis improvement that will be an integral part of the custom designed software platform for processing data collected with Dose-3D phantom. Apart from the innovative detection system the software itself may prove to be disruptive in the context of the currently available tools by offering open-source high quality toolkit for wide use in everyday clinical applications.
ProTheRaMon - a GATE simulation framework for proton therapy range monitoring using PET imaging
D. Borys, J. Baran, K.W. Brzezinski, J. Gajewski, N. Chug, A. Coussat, E. Czerwiński, M. Dadgar, K. Dulski, K. Valsan Eliyan, A. Gajos, K. Kacprzak, Ł. Kapłon, K. Klimaszewski, P. Konieczka, R. Kopec, G. Korcyl, T. Kozik, W. Krzemień, D. Kumar, A. John Lomax, K. McNamara, S. Niedźwiecki, P. Olko, D. Panek, S. Parzych, E. Pérez del Río, L. Raczyński, S. Sharma, S. Shivani, R.Y. Shopa, T. Skóra, M. Skurzok, P. Stasica, E. Stępień, K. Tayefi Ardebili, F. Tayefi, D. Charles Weber, C. Winterhalter, W. Wiślicki, P. Moskal, A. Rucinski
abstract
Objective: This paper reports on the implementation and shows examples of the use of the ProTheRaMon framework for simulating the delivery of proton therapy treatment plans and range monitoring using positron emission tomography (PET). ProTheRaMon offers complete processing of proton therapy treatment plans, patient CT geometries, and intra-treatment PET imaging, taking into account therapy and imaging coordinate systems and activity decay during the PET imaging protocol specific to a given proton therapy facility. We present the ProTheRaMon framework and illustrate its potential use case and data processing steps for a patient treated at the Cyclotron Centre Bronowice (CCB) proton therapy center in Krakow, Poland. Approach: The ProTheRaMon framework is based on GATE Monte Carlo software, the CASToR reconstruction package and in-house developed Python and bash scripts. The framework consists of five separated simulation and data processing steps, that can be further optimized according to the user's needs and specific settings of a given proton therapy facility and PET scanner design. Main results: ProTheRaMon is presented using example data from a patient treated at CCB and the J-PET scanner to demonstrate the application of the framework for proton therapy range monitoring. The output of each simulation and data processing stage is described and visualized. Significance: We demonstrate that the ProTheRaMon simulation platform is a high-performance tool, capable of running on a computational cluster and suitable for multi-parameter studies, with databases consisting of large number of patients, as well as different PET scanner geometries and settings for range monitoring in a clinical environment. Due to its modular structure, the ProTheRaMon framework can be adjusted for different proton therapy centers and/or different PET detector geometries. It is available to the community via github.
Investigation of the light output of 3D-printed plastic scintillators for dosimetry applications
Ł. Kapłon, D. Kulig, S. Beddar, T. Fiutowski, W. Górska, J. Hajduga, P. Jurgielewicz, D. Kabat, K. Kalecińska, M. Kopeć, S. Koperny, B. Mindur, J. Moroń, G. Moskal, S. Niedźwiecki, M. Silarski, F. Sobczuk, T. Szumlak, A. Ruciński
abstract
Three-dimensional (3D) printing, specifically digital light processing (DLP) technique, can be used to manufacture plastic scintillators of any shape. The purpose of this study was to determine the light output of DLP 3Dprinted scintillators for dosimetry applications. Two types of plastic scintillators with dimensions 10 mm × 10 mm × 10 mm were fabricated using DLP 3D-printing at Hanyang University, South Korea. The light output of these DLP 3D-printed samples was measured and compared to that of a commercial plastic scintillator of the same dimensions, RP-408, produced by casting. The 3D-printed scintillators emitting violet and blue light had a lower relative light output by 49% and 43%, respectively, compared to the RP-408 reference scintillator. We also investigated three types of scintillator surface finishing methods: the original surface made by the 3D printer, a sanded surface, and a polished surface. Furthermore, three wrapping configurations were tested: bare scintillator, diffuse-type polytetrafluoroethylene tape, and specular-type enhanced specular reflector foil. Both reflector types, diffuse and specular, reflected blue light with comparable efficiency. Additionally, emission and transmission spectra of the samples were measured. Emission maxima were located at 430 nm for RP-408, and 438 and 475 nm for two 3D-printed samples. Transmittance at the wavelength of maximum emission was equal to 89% for RP-408, and 73% and 66% for the two DLP-printed samples. Although the light output of the 3D-printed scintillators was about 50% lower than that of the commercial plastic scintillator, due to characteristics of 3Dprinted plastic scintillators, i.e. fast, low-cost production, and easy customization of the printed shape, they are promising as an active part of dosimeters for use in high intensity gamma radiation fields produced by medical linear accelerators with acceptable signal-to-noise ratio level.
Positronium imaging with the novel multiphoton PET scanner
P. Moskal, K. Dulski, N. Chug, C. Curceanu, E. Czerwiński, M. Dadgar, J. Gajewski, A. Gajos, G. Grudzień, B.C. Hiesmayr, K. Kacprzak, Ł. Kapłon, H. Karimi, K. Klimaszewski, G. Korcyl, P. Kowalski, T. Kozik, N. Krawczyk, W. Krzemień, E. Kubicz, P. Małczak, S. Niedźwiecki, M. Pawlik-Niedźwiecka, M. Pędziwiatr, L. Raczyński, J. Raj, A. Ruciński, S. Sharma, Shivani, R.Y. Shopa, M. Silarski, M. Skurzok, E.Ł. Stępień, M. Szczepanek, F. Tayefi, W. Wiślicki
abstract
In vivo assessment of cancer and precise location of altered tissues at initial stages of molecular disorders are important diagnostic challenges. Positronium is copiously formed in the free molecular spaces in the patient?s body during positron emission tomography (PET). The positronium properties vary according to the size of inter- and intramolecular voids and the concentration of molecules in them such as, e.g., molecular oxygen, O2; therefore, positronium imaging may provide information about disease progression during the initial stages of molecular alterations. Current PET systems do not allow acquisition of positronium images. This study presents a new method that enables positronium imaging by simultaneous registration of annihilation photons and deexcitation photons from pharmaceuticals labeled with radionuclides. The first positronium imaging of a phantom built from cardiac myxoma and adipose tissue is demonstrated. It is anticipated that positronium imaging will substantially
enhance the specificity of PET diagnostics.
The J-PET detector - a tool for precision studies of ortho-positronium decays
K. Dulski, S.D. Bass, J. Chhokar, N. Chug, C. Curceanu, E. Czerwiński, M. Dadgar, J. Gajewski, A. Gajos, M. Gorgol, R. Del Grande, B.C. Hiesmayr, B. Jasińska, K. Kacprzak, Ł. Kapłon, H. Karimi, D. Kisielewska, K. Klimaszewski, P. Kopka, G. Korcyl, P. Kowalski, T. Kozik, N. Krawczyk, W. Krzemień, E. Kubicz, P. Małczak, M. Mohammed, Sz. Niedźwiecki, M. Pałka, M. Pawlik-Niedźwiecka, M. Pędziwiatr, L. Raczyński7, J. Raj, A. Ruciński, S. Sharma, Shivani, R.Y. Shopa, M. Silarski, M. Skurzok, E. Ł. Stępień, F. Tayefi, W. Wiślicki, B. Zgardzińska, P. Moskal
abstract
The J-PET tomograph is constructed from plastic scintillator strips arranged axially in concentric cylindrical layers. It enables investigations of positronium decays by measurement of the time, position, polarization and energy deposited by photons in the scintillators, in contrast to studies conducted so far with crystal and semiconductor based detection systems where the key selection of events is based on the measurement of the photons energies. In this article we show that the J-PET tomography system constructed solely from plastic scintillator detectors is capable of exclusive measurements of the decays of ortho-positronium atoms. We present the first positronium production results and its lifetime distribution measurements. The obtained results prove the capability of the J-PET tomograph for (i) fundamental studies of positronium decays (in particular test of discrete symmetries in purely leptonic systems), (ii) positron annihilation lifetime spectroscopy, as well as (iii) molecular imaging diagnostics and (iv) observation of entanglement
3D TOF-PET image reconstruction using total variation regularization
L. Raczyński, W. Wiślicki, K. Klimaszewski, W. Krzemień, P. Kopka, P. Kowalski, R. Y. Shopa, M. Bała, J. Chhokar, C. Curceanu, E. Czerwinski, K. Dulski, J. Gajewski, A. Gajos, M. Gorgol, R. Del Grande, B. Hiesmayr, B. Jasińska, K. Kacprzak, L. Kapłon, D. Kisielewska, G. Korcyl, T. Kozik, N. Krawczyk, E. Kubicz, M. Mohammed, S. Niedźwiecki, M. Pałka, M. Pawlik-Niedźwiecka, J. Raj, K. Rakoczy, A. Ruciński, S. Sharma, S. Shivani, M. Silarski, M. Skurzok, E.Ł. Stepień, B. Zgardzińska, P. Moskal
abstract
In this paper we introduce a semi-analytic algorithm for 3-dimensional image reconstruction for positron emission tomography (PET). The method consists of the back-projection of the acquired data into the most likely image voxel according to time-of-flight (TOF) information, followed by the filtering step in the image space using an iterative optimization algorithm with a total variation (TV) regularization. TV regularization in image space is more computationally efficient than usual iterative optimization methods for PET reconstruction with a full system matrix that uses TV regularization. The efficiency comes from the one-time TOF back-projection step that might also be described as a reformatting of the acquired data. An important aspect of our work concerns the evaluation of the filter operator of the linear transform mapping an original radioactive tracer distribution into the TOF back-projected image. We obtain concise, closed-form analytical formula for the filter operator. The proposed method is validated with the Monte Carlo simulations of the NEMA IEC phantom using a one-layer, 50 cm-long cylindrical device called Jagiellonian PET scanner. The results show a better image quality compared with the reference TOF maximum likelihood expectation maximization algorithm.
A simple approach for experimental characterization and validation of proton pencil beam profiles
P. Stasica, J. Baran, C. Granja, N. Krah, G. Korcyl, C. Oancea, M. Pawlik-Niedźwiecka, Sz. Niedźwiecki, M. Rydygier, A. Schavi, A. Rucinski, J. Gajewski
abstract
A precise characterization of therapeutic proton pencil beams is essential for commissioning of any treatment planning system (TPS). The dose profile characterization includes measurement of the beam lateral dose profile in the beam core and far from the beam core, in the so called low-dose envelope, and requires a sophisticated detection system with a few orders of magnitude dynamic range. We propose to use a single-quantum sensitive MINIPIX TIMEPIX detector, along with an in-house designed holder to perform measurements of the pencil beam dose profile in air and in water. We validated the manufacturer calibration of the MINIPIX TIMEPIX detector in proton beams of various energies and compared the deposited energy spectra to Monte Carlo (MC) simulations. The precision of the lateral dose profile measurements has been systematically validated against Krakow proton facility commissioning data and dose profile simulations performed with MC codes GATE/Geant4 and FRED. We obtained an excellent agreement between MINIPIX TIMEPIX measurements and simulations demonstrating the feasibility of the system for a simple characterization and validation of proton pencil beams. The proposed approach can be implemented at any proton therapy facility to acquire experimental data needed to commission and validate analytical and MC based TPS.
Performance assessment of the 2gamma positronium imaging with the total-body PET scanners
P. Moskal, D. Kisielewska, Z. Bura, C. Chhokar, C. Curceanu, E. Czerwiński, M. Dadgar, K. Dulski, J. Gajewski, A. Gajos, M. Gorgol, R. Del Grande, B. C. Hiesmayr, B. Jasińska, K. Kacprzak, A. Kamińska, Ł. Kapłon, H. Karimi, G. Korcyl, P. Kowalski, N. Krawczyk, W. Krzemień, T. Kozik, E. Kubicz, P. Małczak, M. Mohammed, Sz. Niedźwiecki, M. Pałka, M. Pawlik-Niedźwiecka, M. Pędziwiatr, L. Raczyński, J. Raj, A. Ruciński, S. Sharma, Shivani, R. Y. Shopa, M. Silarski, M. Skurzok, E. Ł. Stępień, S. Vandenberghe, W. Wiślicki, B. Zgardzińska
abstract
In living organisms the positron-electron annihilation (occurring during the PET imaging) proceeds in about 30% via creation of a metastable ortho-positronium atom. In the tissue, due to the pick-off and conversion processes, over 98% of ortho-positronia annihilate into two 511~keV photons. In this article we assess the feasibility for reconstruction of the mean ortho-positronium lifetime image based on annihilations into two photons. The main objectives of this work include: (i) estimation of the sensitivity of the total-body PET scanners for the ortho-positronium mean lifetime imaging using 2gamma annihilations, and (ii) estimation of the spatial and time resolution of the ortho-positronium image as a function of the coincidence resolving time (CRT) of the scanner. Simulations are conducted assuming that radiopharmaceutical is labelled with 44Sc isotope emitting one positron and one prompt gamma. The image is reconstructed on the basis of triple coincidence events. The ortho-positronium lifetime spectrum is determined for each voxel of the image. Calculations were performed for cases of total-body detectors build of (i) LYSO scintillators as used in the EXPLORER PET, and (ii) plastic scintillators as anticipated for the cost-effective total-body J-PET scanner. To assess the spatial and time resolution the three cases were considered assuming that CRT is equal to 140ps, 50ps and 10ps. The estimated total-body PET sensitivity for the registration and selection of image forming triple coincidences is larger by a factor of 12.2 (for LYSO PET) and by factor of 4.7 (for plastic PET) with respect to the sensitivity for the standard 2gamma imaging by LYSO PET scanners with AFOV=20cm.
Estimating relationship between the Time Over Threshold and energy loss by photons in plastic scintillators used in the J-PET scanner
S. Sharma, J. Chhokar, C. Curceanu, E. Czerwinski, M. Dadgar, K. Dulski, J. Gajewski, A. Gajos, M. Gorgol, N. Gupta-Sharma, R. Del Grande, B. C. Hiesmayr, B. Jasinska, K. Kacprzak, L. Kaplon, H. Karimi, D. Kisielewska, K. Klimaszewski, G. Korcyl, P. Kowalski, T. Kozik, N. Krawczyk, W. Krzemien, E. Kubicz, M. Mohammed, Sz. Niedzwiecki, M. Palka, M. Pawlik-Niedzwiecka, L. Raczynski, J. Raj, A. Rucinski, Shivani, R. Y. Shopa, M. Silarski, M. Skurzok, E. L. Stepien, W. Wislicki, B. Zgardzinska, P. Moskal
abstract
Time-Over-Threshold (TOT) technique is being used widely due to its implications in developing the multi channel readouts mainly when fast signal processing is required. Using TOT technique as a measure of energy loss instead of charge integration methods significantly reduces the signals readout cost by combining the time and energy information. Therefore, this approach can potentially be used in J-PET tomograph which is build from plastic scintillators characterized by fast light signals. The drawback in adopting this technique is lying in the non-linear correlation between input energy loss and TOT of the signal. The main motivation behind this work is to develop the relationship between TOT and energy loss and validate it with the J-PET tomograph.
The experiment was performed using the 22Na beta emitter source placed in the center of the J-PET tomograph. One can obtain primary photons of two different energies: 511 keV photon from the annihilation of positron (direct annihilation or through the formation of para-Positronim atom or pick-off process of ortho-Positronium atoms), and 1275 keV prompt photon. This allows to study the correlation between TOT values and energy loss for energy range up to 1000 keV. As the photon interacts dominantly via Compton scattering inside the plastic scintillator, there is no direct information of primary photon energy. However, using the J-PET geometry one can measure the scattering angle of the interacting photon. Since, 22Na source emits photons of two different energies, it is required to know unambiguously the energy of incident photons and its corresponding scattering angle for the estimation of energy deposition. In this work, the relationship between Time Over Threshold and energy loss by interacting photons inside the plastic scintillators used in J-PET scanner is established for a energy deposited range 100-1000 keV.
Investigations on physical and biological range uncertainties in Krakow proton beam therapy centre
A. Rucinski, J. Baran, G. Battistoni, A. Chrostowska, M. Durante, J. Gajewski, M. Garbacz, K. Kisielewicz, N. Krah, V. Patera, M. Pawlik-Niedźwiecka, I. Rinaldi, B. Rozwadowska-Bogusz, E. Scifoni, A. Skrzypek, F. Tommasino, A. Schiavi, P. Moskal
abstract
Physical and biological range uncertainties limit the clinical potential of Proton Beam Therapy (PBT). In this proceedings, we report on two research projects, which we are conducting in parallel and which both tackle the problem of range uncertainties. One aims at developing software tools and the other at developing detector instrumentation. Regarding the first, we report on our development and pre-clinical application of a GPU-accelerated Monte Carlo (MC) simulation toolkit Fred. Concerning the letter, we report on our investigations of plastic scintillator based PET detectors for particle therapy delivery monitoring. We study the feasibility of Jagiellonian-PET detector technology for proton beam therapy range monitoring by means of MC simulations of the beta+ activity induced in a phantom by proton beams and present preliminary results of PET image reconstruction. Using a GPU-accelerated Monte Carlo simulation toolkit Fred and plastic scintillator based PET detectors we aim to improve patient treatment quality with protons.