Feasibility of the J-PET to monitor range oftherapeutic proton beams
J. Baran, D. Borys, K. Brzeziński, J. Gajewski, M. Silarski, N. Chug, A. Coussat, E. Czerwiński, M. Dadgar, K. Dulski, K.V. Eliyan, A. Gajos, K. Kacprzak, Ł. Kapłon, K. Klimaszewski, P. Konieczka, R. Kopeć, G. Korcyl, T. Kozik, W. Krzemień, D. Kumar, A.J. Lomax, K. McNamara, S. Niedźwiecki, P. Olko, D. Panek, S. Parzych, E. Perez del Rio, L. Raczyński, M. Simbarashe, S. Sharma, Shivani, R.Y. Shopa, T. Skóra, M. Skurzok, P. Stasica, E.Ł. Stępień, K. Tayefi, F. Tayefi, D.C. Weber, C. Winterhalter, W. Wiślicki, P. Moskal, A. Ruciński
abstract
Objective: The aim of this work is to investigate the feasibility of the JagiellonianPositron Emission Tomography (J-PET) scanner for intra-treatment proton beamrange monitoring. Approach: The Monte Carlo simulation studies with GATE and PET imagereconstruction with CASToR were performed in order to compare six J-PET scannergeometries (three dual-heads and three cylindrical). We simulated proton irradiationof a PMMA phantom with a Single Pencil Beam (SPB) and Spread-Out BraggPeak (SOBP) of various ranges. The sensitivity and precision of each scanner werecalculated, and considering the setup?s cost-effectiveness, we indicated potentiallyoptimal geometries for the J-PET scanner prototype dedicated to the proton beamrange assessment. Main results: The investigations indicate that the double-layer cylindrical andtriple-layer double-head configurations are the most promising for clinical application.We found that the scanner sensitivity is of the order of 10?5coincidences per primaryproton, while the precision of the range assessment for both SPB and SOBP irradiationplans was found below 1 mm. Among the scanners with the same number of detectormodules, the best results are found for the triple-layer dual-head geometry. Significance: We performed simulation studies demonstrating that the feasibilityof the J-PET detector for PET-based proton beam therapy range monitoring ispossible with reasonable sensitivity and precision enabling its pre-clinical tests in theclinical proton therapy environment. Considering the sensitivity, precision and cost-effectiveness, the double-layer cylindrical and triple-layer dual-head J-PET geometryconfigurations seem promising for the future clinical application. Experimental testsare needed to confirm these findings.